Analisis Akurasi dan Pengujian Repeatability Pengukuran Flowmeter Gas pada Laju Aliran Rendah terhadap Variasi Nozzle Meter Gas Diafragma Tipe G1.6
Rizky Candraditya - Dr. Hadi Nasbey, S.Pd., M.Si. - Widya Rachmasari, S. Si., M.T.

Universitas Negeri Jakarta


Abstract

The diaphragm gas meter is a volumetric measuring device commonly used to measure household gas flow at low flow rates. Its reliability depends on measurement accuracy and stability. The Metrology Management Unit of DKI Jakarta is responsible for its testing. This report presents internship research on the accuracy and repeatability of the G1.6 gas meter at low flow rates, using the BPG SONICAL SN G4 device and CalWin SONICAL software at four flow points: 0.016, 0.025, 0.040, and 0.500 cubic meters per hour. Nine units were tested with three repetitions at each point. Results showed that at 0.016, many units displayed under-reading beyond the 3 percent tolerance limit, and measurement stability decreased at lower flow rates. These findings support the development of better calibration procedures and improved accuracy in legal metrology and residential gas services.

Keywords: Diaphragm gas meter, Low flow rate, Nozzle, Metrology

Topic: Instrumentation and Computational Physics

Plain Format | Corresponding Author (Rizky Candraditya)

Share Link

Share your abstract link to your social media or profile page

IPS 2025 - Conference Management System

Powered By Konfrenzi Ultimate 1.832M-Build8 © 2007-2025 All Rights Reserved